JUMO diraVIEW 105 (96 x 96 mm, 1 analog input, 2 relays, 1 digital input, 1 digital input/output)
€354.90*
Please log in to receive your conditions.
Available in 28 days, delivery time 1-2 days
(Login for quotation) The entered number of pieces will be taken into account when creating the quotation.
Product information "JUMO diraVIEW 105 (96 x 96 mm, 1 analog input, 2 relays, 1 digital input, 1 digital input/output)"
Multifunctional digital indicator JUMO diraVIEW 105 for processing and visualization of process signals
JUMO diraVIEW 105 – Digital indicator (96 × 96 mm)
The JUMO diraVIEW 105 is a multifunctional industrial digital indicator for processing and visualization of process signals with extended control and evaluation functions. It is suitable for applications where measured values are to be directly recorded, processed, and integrated into control processes.
The device is a versatile digital indicator for industrial measurement and monitoring tasks and is used for the acquisition, display, and further processing of process signals in machinery and plant environments. It is designed for installation in control panels in a 96 × 96 mm format and enables flexible processing of a wide variety of measured variables.
The device features a universal analog input as well as freely configurable digital inputs and outputs, allowing different sensor signals and control functions to be flexibly combined. This is complemented by relay and logic outputs that enable direct control of external components.
Optional interfaces such as USB, RS485, or Ethernet allow the indicator to be integrated into higher-level systems or conveniently configured. This makes it suitable for both simple display tasks and more complex automation processes.
Functions such as limit value monitoring, timer, min./max. value storage, tare and hold functions, as well as mathematical and logical link functions expand the range of applications and enable direct signal processing within the device. In addition, customer-specific linearizations and individual applications can be implemented.
Advantages
- Universally applicable for different process signals
- Integrated alarm and limit value functions
- Expandable through various communication interfaces
- Direct signal processing possible without external control
- High adaptability through software and logic functions
- Suitable for complex measurement and control tasks
Technical data
- Type: 701514/8-0000-23/000
- Format: 96 × 96 mm
- Input: 1 × universal analog input
- Outputs: 2 × relays, 1 × digital input, 1 × DI/DO (logic 0/14 V)
- Power supply: AC 110–240 V, 48–63 Hz
- Display: LCD with 18-segment and matrix display
- Interfaces: optional USB, RS485, Ethernet
- Functions: Min./max., hold, tare, timer, limit value monitoring
- Type addition: without
Typical areas of application
- Mechanical and plant engineering
- Process and procedural engineering
- Industrial automation
- Test benches and test systems
- Complex measurement and control applications
Dimensions
| Height | 8,6 cm |
|---|---|
| Length | 10,6 cm |
| Width | 9,8 cm |
Weight
| Weight | 286 g |
|---|
Other attributes
| Country of origin | DE |
|---|---|
| Customs tariff number | 90303370 |
| Format | 96 x 96 mm |
| Spannungsversorgung | AC 110 bis 240 V + 10/-15% , 48 bis 63Hz |
Downloads:
| Designation | Type | standard | ITS | Measuring range | Accuracya |
|---|---|---|---|---|---|
| Fe-CuNi | „L“ | DIN 43710 (1985) | IPTS-68 | -200 to +900 °C | ≤ 0,25 % |
| Fe-CuNi | „J“ |
DIN EN 60584-1:2014 IEC 60584-1:2013 |
ITS-90 | -210 to +1200 °C | ≤ 0,25 % from -100 °C |
| Cu-CuNi | „U“ | DIN 43710 (1985) | IPTS-68 | -200 to +600 °C | ≤ 0,25 % from -100 °C |
| Cu-CuNi | „T“ |
DIN EN 60584-1:2014 IEC 60584-1:2013 |
ITS-90 | -270 to +400 °C | ≤ 0,25 % from -150 °C |
| NiCr-Ni | „K“ |
DIN EN 60584-1:2014 IEC 60584-1:2013 |
ITS-90 | -270 to +1300 °C | ≤ 0,25 % from -80 °C |
| NiCr-CuNi | „E“ |
DIN EN 60584-1:2014 IEC 60584-1:2013 |
ITS-90 | -270 to +1000 °C | ≤ 0,25 % from -80 °C |
| NiCrSi-NiSi | „N“ |
DIN EN 60584-1:2014 IEC 60584-1:2013 |
ITS-90 | -270 to +1300 °C | ≤ 0,25 % from -80 °C |
| Pt10Rh-Pt | „S“ |
DIN EN 60584-1:2014 IEC 60584-1:2013 |
ITS-90 | -50 to +1768 °C | ≤ 0,4 % from 20 °C |
| Pt13Rh-Pt | „R“ |
DIN EN 60584-1:2014 IEC 60584-1:2013 |
ITS-90 | -50 to +1768 °C | ≤ 0,25 % from 50 °C |
| Pt30Rh-Pt6Rh | „B“ |
DIN EN 60584-1:2014 IEC 60584-1:2013 |
ITS-90 | -50 to +1820 °C | ≤ 0,25 % from 400 °C |
| W5Re-W26Re | „C“ |
DIN EN 60584-1:2014 IEC 60584-1:2013 |
ITS-90 | 0 to 2315 °C | ≤ 0,25 % from 500 °C |
| W3Re-W25Re | „D“ | ASTM E1751M-15 | ITS-90 | 0 to 2315 °C | ≤ 0,25 % from 500 °C |
| W5Re-W20Re | „A1“ | GOST R 8.585-2001 | ITS-90 | 0 to 2500 °C | ≤ 0,25 % from 500 °C |
| Chromel®-Copel | „L“ | GOST R 8.585-2001 | ITS-90 | -200 to +800 °C | ≤ 0,25 % from -80 °C |
| Chromel®-Alumel® | „K“ | GOST R 8.585-2001 | ITS-90 | -270 to +1300 °C | ≤ 0,25 % from -80 °C |
| a The accuracy applies to the measurement range. | |||||
| Parameter | Description |
|---|---|
| Effect of ambient temperature | ≤ 100 ppm/K |
| Comparison site | internal or external (constant) |
| Reference point temperature (external) | 0 to 100 °C (adjustable) |
| sampling cycle | 150 ms |
| Input filter |
second-order digital filter; Filter constant adjustable from 0 to 100.0 s |
| Designation | standard | ITS | Connection type | Measuring range | Accuracya | Measuring current |
|---|---|---|---|---|---|---|
| Pt100 |
DIN EN 60751:2009 IEC 60751:2008 |
ITS-90 | 2-wire connection | -200 to +850 °C | ≤ 0,2 % | 500 µA |
| 3-wire connection | -200 to +850 °C | ≤ 0,1 % | 500 µA | |||
| Pt1000 |
DIN EN 60751:2009 IEC 60751:2008 |
ITS-90 | 2-/3-wire connection | -200 to +850 °C | ≤ 0,1 % | 50 µA |
| Pt100 | GOST 6651-2009 A.2 | ITS-90 | 2-wire connection | -200 to +850 °C | ≤ 0,2 % | 500 µA |
| 3-wire connection | -200 to +850 °C | ≤ 0,1 % | 500 µA | |||
| KTY | – | – | 2-wire connection | -53 to +153 °C | ≤ 2,0 % | 50 µA |
| a The accuracy applies to the measurement range. | ||||||
| Parameter | Description |
|---|---|
| Effect of ambient temperature | ≤ 50 ppm/K |
| Sensor lead resistance | max. 30 Ω per line |
| sampling cycle | 150 ms |
| Input filter |
second-order digital filter; Filter constant adjustable from 0 to 100.0 s |
| Designation | Measuring range | Accuracya | Measuring current |
|---|---|---|---|
| resistance potentiometer / WFG | 0 to 4000 Ω | ≤ 0,1 % | 50 µA |
| Resistor / potentiometer | 0 to 400 Ω | ≤ 0,1 % | 500 µA |
| 0 to 4000 Ω | ≤ 0,1 % | 50 µA | |
| a The accuracy refers to the maximum measuring range. For small measuring ranges, the linearity accuracy is reduced. | |||
| Parameter | Description |
|---|---|
| Effect of ambient temperature | ≤ 100 ppm/K |
| Connection type |
Resistance potentiometer / WFG: three-wire connection Resistor / potentiometer: two-/three-wire connection |
| Sensor lead resistance | max. 30 Ω per line |
| sampling cycle | 150 ms |
| Input filter |
second-order digital filter; Filter constant adjustable from 0 to 100.0 s |
| Description | Measurement range | accuracya | Input resistance or load voltage |
|---|---|---|---|
| Voltage | 0 to 10 V | ≤ 0,1 % | > 500 kΩ |
| 2 to 10 V | ≤ 0,1 % | > 500 kΩ | |
| Electricity | 4 to 20 mA | ≤ 0,1 % | < 2,5 V |
| 0 to 20 mA | ≤ 0,1 % | < 2,5 V | |
| a The accuracy refers to the maximum measuring range. For small measuring ranges, the linearity accuracy is reduced. | |||
| Parameter | Description |
|---|---|
| The effect of ambient temperature | ≤ 100 ppm/K |
| Measurement range under/over-range |
in accordance with NAMUR recommendation NE 43 (4–20 mA current input only) |
| sampling cycle | 150 ms |
| Input filter |
second-order digital filter; Filter constant adjustable from 0 to 100.0 s |
| measuring transducer | Value below the measurement range | Exceeding the measurement range | Short circuit (sensor/cable) | Break (sensor/cable) | Reverse polarity |
|---|---|---|---|---|---|
| Resistance thermometer | ++ | ++ | ++ | ++ | --- |
| resistor / potentiometer | --- | ++ | --- | ++ | --- |
| Potentiometer / WFG | --- | ++ | (+)a | (+)b | --- |
| thermocouple | ++ | ++ | --- | ++ | (+)c |
| Current 0 to 20 mA | --- | ++ | --- | --- | --- |
| Current 4 to 20 mA | ++ | ++ | ++ | ++ | ++ |
| voltage 0 to 10 V | --- | ++ | --- | --- | ++ |
| voltage 2 to 10 V | ++ | ++ | ++ | ++ | ++ |
|
++ = is recognised |
--- = is not recognised |
(+) = is recognised conditionally a is not recognised in all combinations b A break in the measurement current path is not detected c depending on the selected characteristic curve |
|||||
| Parameter | Description |
|---|---|
| input for volt-free contact |
Contact closed: input is active (RON < 1 kΩ) Contact open: input inactive (ROFF > 50 kΩ) |
| sampling cycle | 150 ms |
| Parameter | Description |
|---|---|
| Voltage – output signal / load resistance | DC 0(2) to 10 V > 500 Ω |
| Current – output signal / load resistance | DC 0(4) to 20 mA < 450 Ω |
| Accuracy | ≤ 0,5 % |
| influence of ambient temperature | ≤ 150 ppm/K |
| output type | switching capacity / output signal | Contact service life |
|---|---|---|
| relay (normally open contact) | max. 3 A at 230 V AC or 30 V DC, resistive load | 150,000 switching operations at rated load |
| – | 350,000 switching operations at 1 A | |
| – | – | |
| relay (normally open contact) with extended contact life | max. 3 A at 230 V AC, resistive load | 300,000 switching operations at rated load |
| – | 1,500,000 switching operations at 1 A | |
| – | – | |
| logic output |
DC 0/14 V ±15 % max. 20 mA per output (at 14 V nominal voltage) |
– |
| PhotoMOS®-Relaisa |
max. 200 mA at 30 V AC or 45 V DC; not short-circuit proof |
– |
| a PhotoMOS is a registered trademark of Panasonic Corporation. | ||
| interface | Parameter | value |
|---|---|---|
| USB-Device | connector type | Micro-B (socket) |
| Standard | Low-Speed, Full-Speed | |
| max. cable length | 5 m | |
| RS485 | baud rate | 9600, 19200, 38400, 115200 |
| data format | 8/1n, 8/1e, 8/1o, 8/2n | |
| protocol | Modbus RTU as slave | |
| Ethernet | connector type | RJ45 (socket) |
| transmission rate | 10 Mbit/s, 100 Mbit/s | |
| protocol |
TCP/IP, DHCP; Modbus-TCP, Modbus-RTU/ASCII over TCP, as Slave |
| Type | Format | Digit height of the upper display | Digit height on the lower display |
|---|---|---|---|
| 701510 | 132 | 7,2 mm | 3,8 mm |
| 701511 | 116 | 12,3 mm | 5,9 mm |
| 701512 | 108H | 11,5 mm | 8,5 mm |
| 701513 | 108Q | 16,5 mm | 9 mm |
| 701514 | 104 | 24,8 mm | 12 mm |
| Parameter | Description |
|---|---|
| Color | top display: white; bottom display: green |
| Figures including decimal places |
top display: 4; lower display: 4 (5 for type 701510, 8 for type 701511) |
| decimal places | 0, 1, 2, 3 or automatically (configurable) |
| Type | Format | Pixel fields |
|---|---|---|
| 701512 | 108H | 2 Rows of 9 pixel fields each |
| 701513 | 108Q | 2 Rows of 8 pixel fields each |
| 701514 | 104 | 2 Rows of 11 pixel fields each |
| Parameter | Value |
|---|---|
| Number of pixels per field | 8 x 5 |
| Color | white |
| Parameter | Variant / Type | Value |
|---|---|---|
| Power supply | Variant 1 | AC 110 to 240 V +10/-15 %, 48 to 63 Hz |
| Variant 2 (see the type plate) | AC/DC 20 to 30 V, 48 to 63 Hz | |
| Electrical safety |
in accordance with DIN EN 61010, Part 1; Surge category II up to 300 V mains voltage, pollution degree 2 |
|
| Protection class | I with internal isolation to SELV | |
| Power consumption | Type 701510 (Format 132) |
at AC 110 to 240 V: max. 3,5 W at AC/DC 20 to 30 V: max. 3,0 W |
| Type 701511 (Format 116) |
at AC 110 to 240 V: max. 4,1 W at AC/DC 20 to 30 V: max. 3,7 W |
|
| Types 701512, 701513 (Formats 108H, 108Q) |
at AC 110 to 240 V: max. 5,8 W at AC/DC 20 to 30 V: max. 5,7 W |
|
| Type 701514 (Format 104) |
at AC 110 to 240 V: max. 6,6 W at AC/DC 20 to 30 V: max. 6,7 W |
|
| Electrical connection | rear via spring-cage terminals (push-in technology) | |
| Conductor cross-section | solid or stranded wire (without ferrule) | min. 0,2 mm², max. 1,5 mm² |
| stranded wire with ferrule |
without plastic collar: min. 0,2 mm², max. 1,5 mm² with plastic collar: min. 0,2 mm², max. 0,75 mm² |
|
| Stripping length | 8 mm | |
| Parameter | Value / classification |
|---|---|
| Ambient temperature range | |
| Storage | -30 to +70 °C |
| Operation | -10 to +55 °C |
| Installation height | max. 2000 m AMSL (above mean sea level) |
| Climatic environmental conditions | in accordance with DIN EN 60721-3, with an extended temperature range |
| climate resistance | ≤ 90 % relative humidity without condensation |
| storage | in accordance with class 1K2 |
| Operation | in accordance with class 3K3 |
| Mechanical environmental conditions | in accordance with DIN EN 60721-3 |
| storage | in accordance with class 1M2 |
| Transport | in accordance with class 2M2 |
| Operation | in accordance with class 3M3 |
| Electromagnetic compatibility (EMC) | in accordance with DIN EN 61326-1 |
| Emitted interference | Class A – for industrial use only – |
| Immunity to interference | industrial requirement |
| Parameter | Type / variant | Value |
|---|---|---|
| Enclosure type | Plastic enclosure for panel mounting in accordance with DIN IEC 61554 (indoor use) | |
| Front panel | made of plastic with membrane keypad | |
| Panel thickness | 1 to 10 mm | |
| Enclosure mounting | in the panel using the supplied mounting frame or the two mounting elements | |
| Operating position | anya | |
| Ingress protection rating | in accordance with DIN EN 60529, front IP65, rear IP20 | |
| Weight | Type 701510 (Format 132) | max. 85 g |
| Type 701511 (Format 116) | max. 120 g | |
| Type 701512 (Format 108H) | max. 160 g | |
| Type 701513 (Format 108Q) | max. 160 g | |
| Type 701514 (Format 104) | max. 220 g | |
| a The maximum permissible ambient temperature applies only when the display is installed in a vertical position. | ||
| Certification marks | Test authority | Certificates / approval numbers | Test basis | Applies to |
|---|---|---|---|---|
| c UL us | Underwriters Laboratories | E201387 |
UL 61010-1 (3. Ed.), CAN/CSA-22.2 No. 61010-1 (3. Ed.) |
all variants |
| DNV GL | DNV GL | TAA00001B3 | Class Guideline DNVGL-CG-0339 |
Type 701511 (DC 20 bis 30 V) and Type 701514 (AC 110 bis 240 V), without Ethernet interface, panel mounting |
| BUREAU VERITAS | Bureau Veritas | 53627/A0 BV | Bureau Veritas Rules for the Classification of Steel Ships | Type 701511 (DC 20 bis 30 V) and Type 701514 (AC 110 bis 240 V), without Ethernet interface, panel mounting |
| The device is approved if the relevant certification mark is displayed on the device. | ||||