| Width | 44 cm |
| Height | 29 cm |
| Length | 56 cm |
| Weight | 11 kg |
| Country of origin | DE |
| Customs tariff number | 90278990 |
Titromat® M1
Price on request
Payment methods
Request individual offer (from 5 pcs.)
For larger order quantities, we will be happy to prepare an individual offer for you.
Delivery & Shipping
- Standard shipping: 3–5 business days within Germany
- Express shipping available on request
- International delivery to EU and Switzerland
- No shipment of Heyl products to Spain, France and Portugal
- Dangerous goods shipping according to ADR regulations
- 3% online discount on HEYL, HORIBA, JUMO and MILWAUKEE products
Order the appropriate original indicator for your TESTOMAT® device:
| Titromat® TC 2010 Reagent A 500 ml | Subscription |
| Titromat® TC 2010 Reagent B 500 ml | Subscription |
Personal consultation
Our experts will be happy to advise you on water treatment, measurement technology and sterile processing.
Technical consulting for industry, laboratory and building technology — your specialist for water analysis and process water treatment.
Contact formQuick info
| Measurement parameters: | Carbonate hardness: CaC03 |
| Measuring range: | 0.89 – 17.9 ppm ; 0.09 - 1.8 °f (resolution: 2 °f) |
| Application: | residual alkalinity after decarbonisation (boiler houses, brewery) |
| Menu navigation: | German, English, French |
| Width | 44 cm |
| Height | 29 cm |
| Length | 56 cm |
| Weight | 11 kg |
| Country of origin | DE |
| Customs tariff number | 90278990 |
Product information "Titromat® M1"
Titromat M1 – Automatic Carbonate Hardness Analysis in the Low Range
Continuous monitoring of water hardness for safe process operations
The Titromat M1 automatically monitors and determines the carbonate hardness in water within the low measurement range. It delivers continuous, precise measurement results and is ideal for systems where fluctuations in water hardness need to be detected immediately.
Technical Features of the Titromat M1
The device provides precise values in the low hardness range and delivers stable measurement data over long periods. Its performance profile includes:
- Flexible measurement intervals: The frequency of measurements can be adjusted to system requirements, allowing rapid changes in water hardness to be detected immediately.
- Automatic sample and reagent handling: The device doses reagents automatically and handles samples on its own, significantly reducing operator effort.
- Documentation and interfaces: Measurement results can be stored directly or transmitted to higher-level systems for process monitoring.
- Long-term stability: Accuracy is maintained even during continuous operation over weeks or months.
Where can the Titromat be used?
- Residual alkalinity after decarbonization (boiler room, brewery)
- Corrosion monitoring in boiler feed water
Why is monitoring carbonate hardness important in water treatment?
Carbonate hardness (residual alkalinity or m-value) significantly affects pH stability, corrosion tendency, and scale formation in water. High values promote deposits and boiler scale, while low values can result in aggressive water and corrosion. Continuous monitoring of carbonate hardness allows targeted process control, protects equipment, and ensures consistent water quality in technical and industrial applications.
Frequently Asked Questions
For monitoring carbonate hardness in the low range. Measuring range: 0.05 - 1 °dH | 0.89 - 17.9 ppm
In boiler systems, breweries (residual alkalinity after decarbonization) and in corrosion control in boiler feed water.
Because even small deviations can have major effects.
It detects fluctuations immediately through the measuring intervals.
Samples and reagents are processed automatically.
Yes, the device delivers consistently accurate values in continuous operation.
Yes, it supports targeted and safe regulation.
| Power consumption: | max. 30 VA |
| Protection class: | I |
| Protection: | IP 65 |
| Dimensions (W x H x D): | 380 x 480 x 280 mm |
| Operating pressure: | 1 bis 8 bar / 1x105 bis 8x105 Pa Oder 0,3* bis 1 bar / 0,3x105 bis 1x105 Pa |
| Weight: | approx. 9,5 kg |
| Menu Language: | German, English, French |
| Measuring range: | 0.89 – 17.9 ppm ; 0.09 - 1.8 °f (resolution: 2 °f) |