Skip to main content Skip to search Skip to main navigation

JUMO contact resistance thermometer with connection cable, protective fitting aluminium (D 10 x 26 mm)

Article number: 802089
Our service hotline – We are here for you 📞 +49 (0) 5121 7609 99

Product information "JUMO contact resistance thermometer with connection cable, protective fitting aluminium (D 10 x 26 mm)"

Surface-Mounted Resistance Thermometer with Pt100 and Tolerance Class B according to DIN 60751 for Temperature Range from -50 to +180 °C

JUMO Surface-Mounted Resistance Thermometer with Connection Cable, Aluminium Protection Fitting (D 10 x 26 mm), Type: 902550/31-378-1003-1-11-2500/000

The surface-mounted resistance thermometer is suitable for indirect temperature measurement on pipes and other surfaces. Equipped with a Pt100 in two-wire circuit, silicone connection cable and a cable length of 2500 mm, this version is particularly suitable for compact surface measurements up to +180 °C.

This surface-mounted resistance thermometer is designed for indirect temperature measurement on surfaces. It is especially suitable for closed piping systems as well as other round or flat surfaces where temperature needs to be measured without interfering with the medium.

Technical Data

  • Design: Surface-mounted resistance thermometer with connection cable
  • Protection fitting: Aluminium, D 10 x 26 mm
  • Sensor insert: 1× Pt100
  • Circuit type: Two-wire circuit
  • Tolerance class: Class B according to DIN EN 60751
  • Connection cable: Silicone
  • Temperature range: -50 to +180 °C
  • Connection cable end: Ferrules according to DIN 46228 Part 4
  • Connection cable length: 2500 mm
  • Type addition: none

Advantages and Technical Features of the JUMO Surface-Mounted Resistance Thermometer

  • indirect temperature measurement without interfering with the system
  • compact aluminium protection fitting D 10 x 26 mm
  • well suited for round and flat surfaces
  • quick and uncomplicated installation
  • low thermal mass for fast temperature response
  • 2500 mm connection cable for flexible installation

The aluminium protection fitting in D 10 x 26 mm makes this version particularly suitable for compact measuring points. Due to the low thermal mass and efficient heat transfer from the measuring point to the sensor, the design is intended to capture surface temperatures quickly and reliably. Especially where installation space is limited, this compact design offers a major advantage.

Temperature measurement is performed using a 1× Pt100 in a two-wire configuration. Combined with the silicone connection cable, this version is designed for temperatures from -50 to +180 °C. This makes the variant suitable for applications requiring a wider temperature range than simpler PVC versions. The 2500 mm long connection cable also provides sufficient flexibility for installation and electrical connection.

Installation is quick and uncomplicated, for example using tension straps or hose clamps. This eliminates the need for mechanical processing of the measuring point. This is particularly useful when retrofitting existing systems or creating an additional measuring point without significant effort.

Since temperature is measured indirectly via the surface, measurement remains unaffected by flow conditions, pressure or chemical influences in the medium. For larger temperature differences between the measured medium and the surrounding environment, additional insulation may be beneficial. Optionally, thermal paste can also be used to further improve heat transfer at the measuring point.

Typical Applications

  • surface temperature measurement on closed piping systems
  • temperature measurement on round surfaces
  • temperature measurement on flat components
  • retrofittable measuring points without mechanical processing

Dimensions

Height 1,3 cm
Length 20 cm
Width 22 cm

Weight

Weight 68 g

Other attributes

Anschlussleitung Silikon
Country of origin DE
Customs tariff number 90251900
Messeinsatz 1x Pt 100 Zweileiterschaltung
Temperaturbereich -20°C ... +80°C
Technical data
Technical specifications Description
Connection Cable ends tinned, available with ferrule terminals, plug-in sleeves or multi-pin connectors
connection cable
  • Silicone, ambient temperature -50 to +180 °C
  • PTFE, ambient temperature -190 to +260 °C
  • Kapton, ambient temperature -50 to +260 °C
Terminal box
  • Basic model 902554/41 and 902554/42: PC plastic housing, colour: silver-grey/blue
  • Basic model 902554/44: PC plastic housing, colour: white
  • Connection via screw terminals; basic model 902554/44 is available with spring-loaded terminals as an option when selected without a transducer
protective tube Stainless steel 1.4571, aluminium, plastic
Gauge insert Pt100 temperature sensor, DIN EN 60751, Class B, two-wire circuit
Transducer Optional for base model 902554/44: analogue transmitter, 4 to 20 mA or 0 to 10 V
Transmitter
Characteristic Output 4 to 20 mA Output 0 to 10 V
Input
Measuring input Pt100 (DIN EN 60751) Pt100 (DIN EN 60751)
Sensor current ≤0,5 mA ≤0,5 mA
Measuring rate Continuous measurement, due to analogue signal path. Continuous measurement, due to analogue signal path.
Measuring circuit monitoring
Underrange (measuring range underrun) decreasing down to ≤ 3,6 mA 0 V
Overrange (measuring range overrun) rising to ≥ 22 up to < 28 mA
(typically 24 mA)
rising to ≥ 11 up to < 14 V
(typically 12 V)
Sensor short circuit ≤ 3,6 mA 0 V
Sensor and cable break ≥ 22 to < 28 mA (typically 24 mA) ≥ 11 to < 14 V (typically 12 V)
Output
Output signal Pulsed direct current 4 to 20 mA DC voltage 0 to 10 V
Transfer behaviour temperature-linear temperature-linear
Transmission accuracy ≤ ±0,1 % ≤ ±0,2 %
Attenuation of the residual ripple of a 24 V power supply, amplitude 10 V / 50 Hz, load 470 Ω / 10 MΩ load 37 dB 40 dB
Load (Rb) Rb = (Ub - 7,5 V) ÷ 22 mA -
Load effect ≤ ±0,02 % / 100 Ωa -
Load / Load Influence - ≥ 10 kΩ / ≤ ±0,1 %
Settling time after temperature change ≤ 10 ms ≤ 10 ms
Calibration conditions DC 24 V / approx. 22 °C DC 24 V / approx. 22 °C
Calibration accuracy ≤ ±0,2 %a,b or ≤ ±0,2 K ≤ ±0,2 %a,b or ≤ ±0,2 K
Overall sensor accuracy/Calibration ±0,4 K (typical) at 20 °C / 24 V
Power supply
±0,4 K (typical) at 20 °C / 24 V
Power supply
Power supply
Power supply (Ub) DC 7,5 at 30 V DC 15 at 30 V
Reverse polarity protection yes yes
Power supply influence ≤ ±0,01 %/V Deviation from 24 Va ≤ ±0,01 %/V Deviation from 24 Va
Environmental influences
Operating temperature range -40 to +85 °C -40 to +85 °C
Lagertemperaturbereich -40 to +100 °C -40 to +100 °C
Temperature influence ≤ ±0,01 %/K deviation from 22 °C a ≤ ±0,01 %/K deviation from 22 °C a
Climatic resistance
similar to DIN EN 60654 class C1
relative humidity ≤ 95% annual average without condensation relative humidity ≤ 95% annual average without condensation
EMC emissions and immunity EN 61326 class B, industrial requirements EN 61326 class B, industrial requirements

a All specifications refer to the full-scale value of 20 mA.
b The higher value applies.

Approvals / certifications
Certification marks Testing body Certificate / test number Basis for audit Applies to
SIL QUALIFIED
PL QUALIFIED
Type suffix 058
in conjunction with the manufacturer's declaration