Typical areas of application
  • industrial waste water technology
  • process measurements, electroplating, paper industry, beverage industry
  • oily waste water
  • suspensions, lacquers, media with solid particles
  • Double-chamber design in the presence of electrode poisons
  • Fluoride-containing (hydrofluoric acid-containing) media up to 1000 mg/l HF

Special features
  • Dirt-repellent PTFE ring diaphragm with highly viscous KCl solution (gel) or perforated or annular gap diaphragm with polymerised solid electrolyte - practically blockage-free
  • Cartridge discharge system with silver ion-free reference electrolyte
  • Pressure-resistant versions up to 10 bar (50 °C)
  • Temperature range: see order details
  • Optional temperature sensors can be integrated
  • Optional salt reservoir to increase service life in media with low conductivity

JUMO order code:

A JUMO order code uniquely identifies a specific measuring instrument or component based on its configuration.

You will find the exact JUMO order code for your instrument configuration in the relevant data sheet from page 5 onwards. This can be found under ‘Downloads’ for the product.

JUMO order code for this product: 201025/51-22-04-22-120/837

Technical data
Active elements of the pH and redox electrode
Membrane glass or active element Designation pH and redox range Temperature range Typical applications
UW glass universal glass pH 0 ... 12
(briefly pH 14)
-5 ... +80 °C water and wastewater engineering,
process instrumentation,
low-ion media
HA-glass high-alkali glass pH 0 ... 14 -5 ... +80 °C In strongly alkaline media (above pH 12)
HT-glass high-temperature glass pH 0 ... 14 0 ... +135 °C At temperatures above 80 °C or
in strongly alkaline media
DS-glass steam-sterilisable glass pH 0 ... 12 -5 ... +80 °C
for short periods up to 130 °C (20 min)
biotechnology, pharmaceutical and food technology,
sterilisation processes
C-glass fluoride-resistant glass pH 0 ... 11 -5 ... +50 °C fluoride-containing (hydrofluoric acid-containing) media
c(HF) ≤ 1000 mg/l
platinum tip redox measurement ±2000 mV -10 ... +135 °C chromate reduction,
nitrite oxidation,
swimming pool and drinking water disinfection
gold tip redox measurement ±2000 mV -10 ... +135 °C cyanide oxidation,
water disinfection
Design variants of the reference system (reference electrode system)
Only silver-ion-free reference electrolytes are used in JUMO tecLine electrodes. A cartridge discharge system contains the silver/silver chloride (Ag/AgCl). Various diaphragm designs are used.
Diaphragm type Description Possible electrolytes Recommended minimum conductivity of the measuring medium Typical applications / limitations
1× Ceramic diaphragm High-quality zirconium dioxide diaphragma Polymerised solid electrolyte Without salt reservoir: 100 µS/cm
With salt reservoir: 50 µS/cm
General water and wastewater engineering,
industrial processes, etc.
liquid KCl (potassium chloride) 5 µS/cm
3× Ceramic diaphragm As above, with increased KCl outflow due to a higher number of outlets Polymerised solid electrolyte Without salt reservoir: 50 µS/cm
With salt reservoir: < 50 µS/cm
In contaminated or low-ion media;
low-temperature applications
liquid KCl (potassium chloride) 0,1 µS/cm
Glass fibre diaphragm Glass fibre bundle instead of a ceramic diaphragm for electrodes with a plastic shaft Polymerised solid electrolyte Without salt reservoir: 150 µS/cm
With salt reservoir: 100 µS/cm
General water and wastewater engineering
(slightly contaminated media)
PTFE ring diaphragm Large-area ring diaphragm Polymerised solid electrolyte Without salt reservoir: 100 µS/cm
With salt reservoir: 50 µS/cm
Only if it is very dirty or e. g. oily, sticky substances
Annular gap or perforated diaphragm Open interface between the solid electrolyte and the medium, arranged in a ring-shaped or point-shaped design Polymerised solid electrolyte Without salt reservoir: 500 µS/cm
With salt reservoir: 500 µS/cm
Suspensions, paints, media containing solids, heavily contaminated media;
not suitable for very pure drinking water or low-ion media

a High-quality ceramic material with consistent porosity, resulting in optimal diffusion properties.