Push-in resistance thermometer with straight connection cable, stainless steel protection tube
Price on request
Product information "Push-in resistance thermometer with straight connection cable, stainless steel protection tube"
JUMO Resistance Thermometer with Insertion Function for Temperature Measurements from -50 to +180 °C
JUMO insertion resistance thermometer with connection cable and stainless steel protection tube, type: 902150/10-378-1003-1-5,2-45-11-2500/000
The insertion resistance thermometer is suitable for precise temperature measurements in liquid and gaseous media. Thanks to its compact design with a 45 mm insertion length, this version is particularly suitable for measuring points where space is limited or where the temperature needs to be recorded close to the installation point.
The sensor is designed with a Pt100 in a two-wire circuit and a 2500 mm long silicone connection cable. The permissible operating range of the connection cable is -50 to +180 °C. This makes this version suitable for many technical applications where reliable temperature measurement is required in dry or humid environments.
Technical Data
- Version: insertion resistance thermometer with straight connection cable
- Protection tube material: stainless steel
- Measuring insert: 1 x Pt100
- Circuit type: two-wire circuit
- Tolerance class: Class B
- Operating temperature / connection cable: -50 to +180 °C / silicone
- Insertion length: 45 mm
- Cable length: 2500 mm
- Protection tube diameter: standard at Ø 5 mm and Ø 5.2 mm
- Connection end: ferrules
- Type suffix: with bend protection, shielded cable
Advantages and Technical Features of the JUMO Insertion Resistance Thermometer
- Compact design for smaller measuring points
- Well suited for measurements close to the installation point
- Flexible silicone connection cable with a length of 2500 mm
- Shielded cable for reliable signal transmission
- Bend protection for improved protection in daily use
- Robust design with stainless steel protection tube
In everyday use, this version offers several practical advantages. The straight connection cable can be routed flexibly, while the ferrules support a clean electrical connection. Bend protection and the shielded cable also help protect the connection during operation against mechanical stress and interference.
The stainless steel protection tube ensures a robust design and makes this version suitable for many typical measuring tasks in technical environments. Especially in applications where compact sensors with a connection cable are required, this version is a practical solution.
Typical Application Areas
- Laboratory technology
- Climate chambers
- Solar systems
- Heating, ventilation and air conditioning
Dimensions
| Height | 1,2 cm |
|---|---|
| Length | 23 cm |
| Width | 19 cm |
Weight
| Weight | 48 g |
|---|
Other attributes
| Country of origin | DE |
|---|---|
| Customs tariff number | 90251900 |
| Einbaulänge in mm | 45 mm |
| Messeinsatz | 1x Pt 100 Zweileiterschaltung |
| Schutzrohrdurchmesser | 5 mm |
| Temperaturbereich | -50°c ... +180°C |
| Electrical connection | Description |
|---|---|
| Connection element |
Connection cable with stripped bare wire end Connection cable with tinned connecting wires Connection cable with wire end ferrules Connection cable with M12 connector Connection cable with multi-pin connector |
| Permissible ambient temperature of the connection cable | Temperature range |
|---|---|
| PVC | -5 to +80 °C |
| Silicone | -50 bis +180 °C |
| PTFE | -50 to +260 °C |
| Metal braid | -50 to +260 °C |
| Glass silk | -50 to +400 °C |
| Merkmal | Description |
|---|---|
|
Materials Insulation material of the connection cable |
PVC Silicone PTFE Metal braid Glass silk |
| protective tube | Stainless steel |
| connection cable | Strain-relieved |
| protective tube |
Ø 3 mm Ø 4 mm Ø 5 mm Ø 6 mm |
|
Response times Iin water 0,4 m/s In the air 3,0 m/s |
t0,5 ≈ 4 s, t0,9 ≈ 10 s, Ø 6 mm t0,5 ≈ 32 s, t0,9 ≈ 98 s, Ø 6 mm |
| Gauge insert |
DIN EN IEC 60751 1× Pt100 in two-wire configuration 1× Pt100 in three-wire configuration 1× Pt100 in four-wire configuration 1× Pt1000 in two-wire configuration 2× Pt100 in two-wire configuration 2× Pt100 in three-wire configuration 2× Pt1000 in two-wire configuration |
| Tolerance class |
Class B (standard) Class A Class AA |