Service kit Professional for 2-year maintenance Testomat® 808 SiO2

Product number: 890612

€1,301.15*

Please log in to receive your conditions.

Available in 28 days, delivery time 1-2 days


Product number: 890612
GTIN: 4068823007527
Our comment: For units with serial number 253.061 and higher, an adjustment with silicate-free water is required after installation. For units with an older serial number, a software update must be carried out in addition to the adjustment mentioned above. Please refer to the operating instructions for this.
Offer as PDF  
Login for addressed offer
The entered number of pieces will be taken into account when creating the offer.
Product information "Service kit Professional for 2-year maintenance Testomat® 808 SiO2"

Service set contains:
2 x sight glasses 30 x 3
3 x flat gasket 24 x 2
1 x insert for screw cap & suction tube 100 ml bottle
1 x insert for screw cap & suction tube 500 ml bottle
1 x double pump head conversion kit
1 x gear motor
2 x Hose connector
1 x Cleaning brush set


Important:
After replacement, the pump must be recalibrated in Testomat® 808 SiO2.
The adjustment is available from firmware version T808_83-003. Carry out a firmware update for older versions before performing the adjustment! (Serial number less than 253.061)
For this calibration, it is important that you use water with a silicate content of 0 ppm, otherwise the zero point calibration will be incorrect and lead to false measurement results.
To ensure this, please use the silicate filter cartridge specially developed for this process for the T808 SiO2, article 270344.
(For additional information and instructions on the equipment you need for the software update, please refer to the operating instructions. The instructions "Replacing the pump head on Testomat® 808 SiO2" are also available)

Dimensions

Weight

Other attributes

Country of origin DE
Customs tariff number 90278990

Required filter cartridge for pump head replacement

Silicate filter cartridge T808 SiO2 for changing the double pump head
Product number: 270344
Dear customers, please note that the calibration of the Testomat® 808 SiO2 can only be carried out with silicate-free water. Therefore, please use the silicate filter cartridge specially developed for this purpose when replacing the double pump head (art. no. 40395) and the associated pump adjustment.The cartridge filled with the special resin is inserted into the sampling line of the unit before the inlet and removed again after the adjustment. The pump replacement instructions describe the use of the silicate filter, storage conditions and shelf life. After the first use, the filter can be used up to 10 times within one year for balancing. After one year, the filter must be disposed of.

€216.22*