Skip to main content Skip to search Skip to main navigation

JUMO outdoor resistance thermometers Protection class IP65

Article number: 802072
Our service hotline – We are here for you 📞 +49 (0) 5121 7609 99

Product information "JUMO outdoor resistance thermometers Protection class IP65"

Jumo Outdoor Resistance Thermometer with Pt1000 in Two-Wire Circuit with Operating Temperature -50 to +90 °C

JUMO Outdoor Resistance Thermometer Protection Class IP65 Type: 902520/15-361-1005-1/903

This resistance thermometer from JUMO is suitable for temperature measurements outdoors and in environments exposed to moisture. Equipped with a Pt1000 sensor in a two-wire circuit, a measuring range of -50 to +90 °C, and an IP65 cable gland, this version is designed for reliable outdoor temperature measurement in HVAC applications.

This version is therefore intended for outdoor temperature measurements where the housing must be reliably protected against dust and moisture.

Technical Data

  • Version: Outdoor resistance thermometer
  • Measuring insert: 1 x Pt1000
  • Circuit type: Two-wire circuit
  • Transmitter: without
  • Operating temperature: -50 to +90 °C
  • Protection class: IP65
  • Special feature: Cable gland (IP65)

Advantages and Technical Features of the JUMO Outdoor Resistance Thermometer

  • well suited for outdoor temperature measurements
  • IP65 protection class for increased housing protection requirements
  • IP65 cable gland for protected cable entry
  • Pt1000 sensor for compatible control and evaluation systems

Unlike Pt100 versions, this model operates with 1 x Pt1000 in a two-wire circuit. This is especially useful when the connected control or evaluation system is designed for Pt1000 or when this sensor configuration is preferred. Since no transmitter is integrated, this version is well suited for applications where the sensor is connected directly to compatible evaluation or control systems.

A practical advantage is the IP65 cable gland. It ensures protected cable entry and supports reliable sealing of the connection point. This is especially important for outdoor applications or technically demanding environments to ensure reliable measurement performance under changing environmental conditions.

Typical Applications

  • Outdoor temperature measurement on buildings
  • Heating, ventilation, and air conditioning systems
  • Technical outdoor areas
  • Weather-exposed measuring points

Dimensions

Height 4 cm
Length 17 cm
Width 15 cm

Weight

Weight 65 g

Other attributes

Country of origin DE
Customs tariff number 90251900
Messeinsatz 1x Pt 1000 Zweileiterschaltung
Schutzart IP65
Temperaturbereich -50°C ... +90°C
Technical Data:
Technical Data
Feature Description
Connection housing Plastic housing made of PC (basic type 902520/11, material: PP), degree of protection IP20 to IP65.
Basic type 902524/25, degree of protection IP54 and IP65.
Protective tube Stainless steel 1.4571; Ø 6 mm
Measuring insert Pt100 temperature sensor, DIN EN 60751, class B, two-wire circuit.
Pt1000 temperature sensor, DIN EN 60751, class B, two-wire circuit.
Further versions, see ordering information.
Transmitter Analogue transmitter, output signal 4 to 20 mA or 0 to 10 V.
Transmitter
Characteristic Output 4 to 20 mA Output 0 to 10 V
Input
Measuring input Pt100 (DIN EN 60751) Pt100 (DIN EN 60751)
Sensor current ≤0,5 mA ≤0,5 mA
Measuring rate Continuous measurement, due to analogue signal path. Continuous measurement, due to analogue signal path.
Measuring circuit monitoring
Underrange (measuring range underrun) decreasing down to ≤ 3,6 mA 0 V
Overrange (measuring range overrun) rising to ≥ 22 up to < 28 mA
(typically 24 mA)
rising to ≥ 11 up to < 14 V
(typically 12 V)
Sensor short circuit ≤ 3,6 mA 0 V
Sensor and cable break ≥ 22 to < 28 mA (typically 24 mA) ≥ 11 to < 14 V (typically 12 V)
Output
Output signal Pulsed direct current 4 to 20 mA DC voltage 0 to 10 V
Transfer behaviour temperature-linear temperature-linear
Transmission accuracy ≤ ±0,1 % ≤ ±0,2 %
Attenuation of the residual ripple of a 24 V power supply, amplitude 10 V / 50 Hz, load 470 Ω / 10 MΩ load 37 dB 40 dB
Load (Rb) Rb = (Ub - 7,5 V) ÷ 22 mA -
Load effect ≤ ±0,02 % / 100 Ωa -
Load / Load Influence - ≥ 10 kΩ / ≤ ±0,1 %
Settling time after temperature change ≤ 10 ms ≤ 10 ms
Calibration conditions DC 24 V / approx. 22 °C DC 24 V / approx. 22 °C
Calibration accuracy ≤ ±0,2 %a,b or ≤ ±0,2 K ≤ ±0,2 %a,b or ≤ ±0,2 K
Overall sensor accuracy/Calibration ±0,4 K (typical) at 20 °C / 24 V
Power supply
±0,4 K (typical) at 20 °C / 24 V
Power supply
Power supply
Power supply (Ub) DC 7,5 at 30 V DC 15 at 30 V
Reverse polarity protection yes yes
Power supply influence ≤ ±0,01 %/V Deviation from 24 Va ≤ ±0,01 %/V Deviation from 24 Va
Environmental influences
Operating temperature range -40 to +85 °C -40 to +85 °C
Lagertemperaturbereich -40 to +100 °C -40 to +100 °C
Temperature influence ≤ ±0,01 %/K deviation from 22 °C a ≤ ±0,01 %/K deviation from 22 °C a
Climatic resistance
similar to DIN EN 60654 class C1
relative humidity ≤ 95% annual average without condensation relative humidity ≤ 95% annual average without condensation
EMC emissions and immunity EN 61326 class B, industrial requirements EN 61326 class B, industrial requirements

a All specifications refer to the full-scale value of 20 mA.
b The higher value applies.

Approvals / certifications
Certification marks Testing body Certificate / test number Basis for audit Applies to
SIL QUALIFIED
PL QUALIFIED
Type suffix 058