Skip to main content Skip to search Skip to main navigation

JUMO room resistance thermometers Protection class IP20

Article number: 802063
Our service hotline – We are here for you 📞 +49 (0) 5121 7609 99

Product information "JUMO room resistance thermometers Protection class IP20"

JUMO room resistance thermometer for temperature measurement in indoor areas and HVAC applications for measuring range -30 to +80 °C

JUMO room resistance thermometer with Pt100 in three-wire configuration and IP20 protection class Type: 902520/10-572-1001-1/00

The room resistance thermometer with IP20 protection class is designed for temperature measurement in indoor areas and HVAC applications. With Pt100 in three-wire configuration and a measuring range from -30 to +80 °C, this version is suitable for reliable room temperature detection.

This room resistance thermometer is designed for temperature measurement in indoor environments and is particularly suitable for HVAC applications. With a Pt100 temperature sensor in three-wire configuration and an operating range from -30 to +80 °C, this version is designed for reliable temperature detection in typical room-related applications.

Technical Data

  • Design: Room resistance thermometer
  • Measuring insert: 1 x Pt100
  • Circuit type: Three-wire configuration
  • Transmitter: none
  • Operating temperature: -30 to +80 °C
  • Protection class: IP20

Advantages and technical features

  • suitable for classic room temperature measurements
  • well suited for HVAC applications
  • three-wire configuration for more accurate temperature measurement
  • suitable for temperatures from -30 to +80 °C

Its strengths are especially apparent wherever temperatures in rooms need to be continuously monitored without requiring a complex sensor design. The IP20 protection class makes this version suitable for dry indoor areas where the sensor is primarily used for room temperature detection. This makes the device well suited for classic building services and HVAC applications.

The integrated Pt100 sensor belongs to the proven solutions in temperature measurement technology. In this version, it is designed as a three-wire configuration, which is particularly advantageous when cable influences should be compensated better than with a simple two-wire configuration. This supports the most precise and stable temperature measurement possible during operation.

Since no transmitter is integrated, this version is particularly suitable for applications where the sensor should be connected directly to compatible evaluation or control technology. This can be useful wherever simple and direct integration into existing systems is required.

Typical Areas of Application

  • room temperature measurement in HVAC applications
  • temperature monitoring in indoor environments
  • building services engineering
  • ventilation and air conditioning systems
  • dry indoor areas with low mechanical stress

Frequently Asked Questions

It reliably measures temperature in indoor spaces and HVAC systems.

Depending on the specific model, the measuring range is −30 °C to +80 °C, 0 °C to +40 °C, or 0 °C to +60 °C.

Yes, it is specifically designed for air conditioning and ventilation applications.

In dry indoor areas such as offices, plant rooms, and general building interiors.

It connects directly to compatible evaluation or control equipment.

As standard, it uses a Pt100 temperature sensor compliant with DIN EN 60751.

It has an IP20 protection rating (protection against solid foreign objects, but not against water).

Dimensions

Height 3 cm
Length 12 cm
Width 8 cm

Weight

Weight 50 g

Other attributes

Country of origin DE
Customs tariff number 90251900
Messeinsatz 1x Pt 100 Dreileiterschaltung
Schutzart IP20
Temperaturbereich -30°C ... +80°C
Toleranz / Klasse Klasse B
Technical data
Technical Data
Feature Description
Connection housing Plastic housing made of PC (basic type 902520/11, material: PP), degree of protection IP20 to IP65.
Basic type 902524/25, degree of protection IP54 and IP65.
Protective tube Stainless steel 1.4571; Ø 6 mm
Measuring insert Pt100 temperature sensor, DIN EN 60751, class B, two-wire circuit.
Pt1000 temperature sensor, DIN EN 60751, class B, two-wire circuit.
Further versions, see ordering information.
Transmitter Analogue transmitter, output signal 4 to 20 mA or 0 to 10 V.
Transmitter
Characteristic Output 4 to 20 mA Output 0 to 10 V
Input
Measuring input Pt100 (DIN EN 60751) Pt100 (DIN EN 60751)
Sensor current ≤0,5 mA ≤0,5 mA
Measuring rate Continuous measurement, due to analogue signal path. Continuous measurement, due to analogue signal path.
Measuring circuit monitoring
Underrange (measuring range underrun) decreasing down to ≤ 3,6 mA 0 V
Overrange (measuring range overrun) rising to ≥ 22 up to < 28 mA
(typically 24 mA)
rising to ≥ 11 up to < 14 V
(typically 12 V)
Sensor short circuit ≤ 3,6 mA 0 V
Sensor and cable break ≥ 22 to < 28 mA (typically 24 mA) ≥ 11 to < 14 V (typically 12 V)
Output
Output signal Pulsed direct current 4 to 20 mA DC voltage 0 to 10 V
Transfer behaviour temperature-linear temperature-linear
Transmission accuracy ≤ ±0,1 % ≤ ±0,2 %
Attenuation of the residual ripple of a 24 V power supply, amplitude 10 V / 50 Hz, load 470 Ω / 10 MΩ load 37 dB 40 dB
Load (Rb) Rb = (Ub - 7,5 V) ÷ 22 mA -
Load effect ≤ ±0,02 % / 100 Ωa -
Load / Load Influence - ≥ 10 kΩ / ≤ ±0,1 %
Settling time after temperature change ≤ 10 ms ≤ 10 ms
Calibration conditions DC 24 V / approx. 22 °C DC 24 V / approx. 22 °C
Calibration accuracy ≤ ±0,2 %a,b or ≤ ±0,2 K ≤ ±0,2 %a,b or ≤ ±0,2 K
Overall sensor accuracy/Calibration ±0,4 K (typical) at 20 °C / 24 V
Power supply
±0,4 K (typical) at 20 °C / 24 V
Power supply
Power supply
Power supply (Ub) DC 7,5 at 30 V DC 15 at 30 V
Reverse polarity protection yes yes
Power supply influence ≤ ±0,01 %/V Deviation from 24 Va ≤ ±0,01 %/V Deviation from 24 Va
Environmental influences
Operating temperature range -40 to +85 °C -40 to +85 °C
Lagertemperaturbereich -40 to +100 °C -40 to +100 °C
Temperature influence ≤ ±0,01 %/K deviation from 22 °C a ≤ ±0,01 %/K deviation from 22 °C a
Climatic resistance
similar to DIN EN 60654 class C1
relative humidity ≤ 95% annual average without condensation relative humidity ≤ 95% annual average without condensation
EMC emissions and immunity EN 61326 class B, industrial requirements EN 61326 class B, industrial requirements

a All specifications refer to the full-scale value of 20 mA.
b The higher value applies.

Approvals / certifications
Certification marks Testing body Certificate / test number Basis for audit Applies to
SIL QUALIFIED
PL QUALIFIED
Type suffix 058
in conjunction with the manufacturer's declaration