JUMO resistance thermometer for room temperature measurement with Pt100 in two-wire configuration for measuring range 0 to +60 °C

JUMO room resistance thermometer IP20 protection class with transmitter 4 to 20 mA Type: 902520/10-807-1003-1/330, (0..60°C)

The room resistance thermometer with IP20 protection class is suitable for room temperature measurement with direct analog signal transmission. With Pt100 in two-wire configuration and integrated 4 to 20 mA transmitter, this version is particularly suitable for indoor HVAC applications and building automation in the range from 0 to +60 °C.

This version is intended for applications where room temperatures should not only be measured but also transmitted directly as an analog signal to a control or regulation system. The room resistance thermometer operates with 1 x Pt100 in two-wire configuration and is additionally equipped with an analog transmitter with 4 to 20 mA output.

Technical Data

  • Design: Room resistance thermometer
  • Measuring insert: 1 x Pt100
  • Circuit type: Two-wire configuration
  • Transmitter: 1 x analog, output 4 to 20 mA
  • Measuring range: 0 to +60 °C
  • Note: Measuring range only in combination with transmitter
  • Protection class: IP20

Advantages and technical features of the JUMO room resistance thermometer

  • well suited for room temperature ranges up to +60 °C
  • no additional external transmitter required
  • easy signal processing within the system
  • suitable for dry indoor environments

The measuring range is from 0 to +60 °C and in this version is linked to the integrated transmitter. This makes the variant particularly suitable for applications where temperatures in rooms or other indoor areas should be continuously monitored and processed as a standard signal. Compared to narrower measuring ranges, this version provides slightly more flexibility at higher temperatures and is therefore also suitable for warmer indoor environments.

With IP20 protection class, the device is intended for dry indoor areas. This makes it well suited for classic room applications in HVAC systems where no higher protection class is required. The integrated transmitter is especially practical where a direct 4 to 20 mA signal is required on the system side and no additional signal conversion is planned.

Typical Areas of Application

  • room temperature measurement in HVAC applications
  • building automation
  • heating, ventilation and air conditioning systems
  • indoor environments with analog signal transmission
  • indoor areas with a slightly extended temperature range

JUMO order code:

A JUMO order code uniquely identifies a specific measuring instrument or component based on its configuration.

You will find the exact JUMO order code for your instrument configuration in the relevant data sheet from page 6 onwards. This can be found under ‘Downloads’ for the product.

Frequently Asked Questions

It reliably measures temperature in indoor spaces and HVAC systems.

Depending on the specific model, the measuring range is −30 °C to +80 °C, 0 °C to +40 °C, or 0 °C to +60 °C.

Yes, it is specifically designed for air conditioning and ventilation applications.

In dry indoor areas such as offices, plant rooms, and general building interiors.

It connects directly to compatible evaluation or control equipment.

As standard, it uses a Pt100 temperature sensor compliant with DIN EN 60751.

It has an IP20 protection rating (protection against solid foreign objects, but not against water).

Dimensions

Height 2 cm
Length 10 cm
Width 8,5 cm

Weight

Weight 54 g

Other attributes

Country of origin DE
Customs tariff number 90251900
Messeinsatz 1x Pt 100 Zweileiterschaltung
Messumformer Messumformer analog Ausgang 4-20 mA
Temperaturbereich 0°C ... +60°C
Toleranz / Klasse Klasse B
Technical data
Technical Data
Feature Description
Connection housing Plastic housing made of PC (basic type 902520/11, material: PP), degree of protection IP20 to IP65.
Basic type 902524/25, degree of protection IP54 and IP65.
Protective tube Stainless steel 1.4571; Ø 6 mm
Measuring insert Pt100 temperature sensor, DIN EN 60751, class B, two-wire circuit.
Pt1000 temperature sensor, DIN EN 60751, class B, two-wire circuit.
Further versions, see ordering information.
Transmitter Analogue transmitter, output signal 4 to 20 mA or 0 to 10 V.
Transmitter
Characteristic Output 4 to 20 mA Output 0 to 10 V
Input
Measuring input Pt100 (DIN EN 60751) Pt100 (DIN EN 60751)
Sensor current ≤0,5 mA ≤0,5 mA
Measuring rate Continuous measurement, due to analogue signal path. Continuous measurement, due to analogue signal path.
Measuring circuit monitoring
Underrange (measuring range underrun) decreasing down to ≤ 3,6 mA 0 V
Overrange (measuring range overrun) rising to ≥ 22 up to < 28 mA
(typically 24 mA)
rising to ≥ 11 up to < 14 V
(typically 12 V)
Sensor short circuit ≤ 3,6 mA 0 V
Sensor and cable break ≥ 22 to < 28 mA (typically 24 mA) ≥ 11 to < 14 V (typically 12 V)
Output
Output signal Pulsed direct current 4 to 20 mA DC voltage 0 to 10 V
Transfer behaviour temperature-linear temperature-linear
Transmission accuracy ≤ ±0,1 % ≤ ±0,2 %
Attenuation of the residual ripple of a 24 V power supply, amplitude 10 V / 50 Hz, load 470 Ω / 10 MΩ load 37 dB 40 dB
Load (Rb) Rb = (Ub - 7,5 V) ÷ 22 mA -
Load effect ≤ ±0,02 % / 100 Ωa -
Load / Load Influence - ≥ 10 kΩ / ≤ ±0,1 %
Settling time after temperature change ≤ 10 ms ≤ 10 ms
Calibration conditions DC 24 V / approx. 22 °C DC 24 V / approx. 22 °C
Calibration accuracy ≤ ±0,2 %a,b or ≤ ±0,2 K ≤ ±0,2 %a,b or ≤ ±0,2 K
Overall sensor accuracy/Calibration ±0,4 K (typical) at 20 °C / 24 V
Power supply
±0,4 K (typical) at 20 °C / 24 V
Power supply
Power supply
Power supply (Ub) DC 7,5 at 30 V DC 15 at 30 V
Reverse polarity protection yes yes
Power supply influence ≤ ±0,01 %/V Deviation from 24 Va ≤ ±0,01 %/V Deviation from 24 Va
Environmental influences
Operating temperature range -40 to +85 °C -40 to +85 °C
Lagertemperaturbereich -40 to +100 °C -40 to +100 °C
Temperature influence ≤ ±0,01 %/K deviation from 22 °C a ≤ ±0,01 %/K deviation from 22 °C a
Climatic resistance
similar to DIN EN 60654 class C1
relative humidity ≤ 95% annual average without condensation relative humidity ≤ 95% annual average without condensation
EMC emissions and immunity EN 61326 class B, industrial requirements EN 61326 class B, industrial requirements

a All specifications refer to the full-scale value of 20 mA.
b The higher value applies.

Approvals / certifications
Certification marks Testing body Certificate / test number Basis for audit Applies to
SIL QUALIFIED
PL QUALIFIED
Type suffix 058
in conjunction with the manufacturer's declaration