Testomat® LAB TH
€3,501.00*
% €3,764.52*Please log in to receive your conditions.
Available in 28 days, delivery time 1-2 days
Register warranty here
A warranty extension is possible after taking out an indicator subscription.
Testomat® indicator TH 2005, 500ml | Order · Subscription |
Testomat® indicator TH 2025 500ml | Order · Subscription |
Testomat® indicator TH2050 500ml | Order · Subscription |
Testomat® indicator TH 2100 500ml. | Order · Subscription |
Testomat® indicator TH 2250 500ml. | Order · Subscription |
(Login for quotation) The entered number of pieces will be taken into account when creating the offer.
Leasing Example
73,70 €
non-binding
Monthly Rate (60 Rates)
Prices excl. VAT
Further rate options by choosing the payment method Leasing
Product information "Testomat® LAB TH"
- Execution of parameterization by SD card or the "Service Monitor" software via mini-USB (on the device side) to USB 2.0 (for example PC/notebook)
- Multilingual menu navigation software in German, English, French, Dutch, Portuguese
- Firmware update via SD card
- compact design
- optimized water detection based on an optical measuring method
- Integrated self-test with continuous monitoring
- Automatic venting of the indicator line
- Parameterisation and readout of the unit parameters via the RS232 interface
- High-precision titration using a piston dosing pump
- Reliable and low-maintenance operation
- Minimal indicator and water consumption
- Analog output 4 - 20 mA for transmission of measured values
- Programmable hardness unit in °dH, °f, ppm CaCO3, mmol/l
- Common alarm output
- Indicator quantities: 500 ml bottles
- Automatic interval operation (interval pause adjustable from 0-255 minutes)
- External analysis input (start/stop)
- Hand launch
- OLED display
- Protecting cover to increase the IP protection class
Dimensions
Height | 21 cm |
---|---|
Length | 43 cm |
Width | 35 cm |
Weight
Weight | 4,55 kg |
---|
Other attributes
Country of origin | EU |
---|---|
Customs tariff number | 90278990 |
Operating voltage: | 24 VDC |
Power consumption: | Max. 1 A, without external load |
Protection class: | I |
Protection: | IP 40 (when using the optionally available device cover, IP43) |
Ambient temperature: | 10–40 °C |
Water temperature: | 10–40 °C |
Dimensions (W x H x D): | 270 x 350 x 147 mm |
Operating pressure: | 1 to 8 bar / 1x105 to 8x105 Pa or 0.3* to 1 bar / 1x105 to 1x105 Pa (*after removing the regulator core) |
Weight: | approx. 4.35 kg |
Menu Language: | (Servicemonitor) German, English, French, Dutch |
Measuring range: | 0.89 - 448 ppm |
Tender documents available for download as Word (*.docx) or PDF (*.pdf) files DE/EN
Tender specification - Testomat® LAB TH
Performance description
Fully automatic titration device for determining the residual total hardness (water hardness) for reliable quality control of drinking, process and heating feed water. Specially designed for use in multi-parameter systems and for integration into higher-level control systems.
Device type
Compact online analyzer based on titration measurement. The device is particularly suitable for use in complex measuring networks and enables seamless integration via interfaces such as 4 - 20 mA and RS 232.
Manufacturer: Gebrüder Heyl Analysentechnik GmbH & Co. KG, Hildesheim
Product name: Testomat® LAB TH
Item number: 116102
Measuring principle & measuring range
Titration by means of color change using a piston metering pump for precise analysis of total hardness in the range of 0.05-25 °dH (or 0.89-448 ppm), depending on the indicator used.
Areas of application
Ideal for monitoring water hardness in the following areas:
- Softening systems
- Water treatment plants
- Drinking water stations
- Industrial boilers
- Process water systems
Technical design
- Automatic sampling and venting of the dosing pump
- Optical water detection for level monitoring and evaluation of sample water quality
- Multilingual menu navigation
- Freely selectable measuring interval times (cyclical or quantity-dependent)
- External analysis triggering
- RS-232 interface (measured value, status, error messages) and analog output 0/4-20 mA
- Storage of all system data and logs (measured values and error messages) on SD card
- Display units: ° dH, ° f, ppm CaCO₃, mmol/l
- Internal memory with data export/import function (SD card)
- Robust housing: IP40 (optionally up to IP 43 with device cover) / protection class I
- Supply voltage: 24 V DC
- current consumption approx. 1 A (without external load)
- Dimensions: approx. 270 x 350 x 147 mm
- Weight: approx. 4,350 grams
- Permissible ambient temperature: 10-40 °C
- Permissible water temperature: 10-40 °C
- Operating pressure: 1-8 bar (or 0.3-1 bar after removing the regulator core)
- Firmware update via SD card
- Parameterization via SD card or service monitor via mini-USB
- Integrated self-test
- Automatic venting of the indicator line
Scope of delivery
- Testomat® LAB TH device
- Integrated interface hardware: 4-20 mA, RS-232, SD card slot
- SD card or SDHC (optional) for data acquisition in CSV format
- Operating and service documentation
Optional accessories and operating materials
- OLED display
- Protective cover (to increase the degree of protection)
- Connection set for sample inlet/outlet consisting of PE hose 6/4x1 for the water inlet, 2 m drain hose di = 12 mm, 1 ball valve, 1 reducing connector 10 to 6 mm and reducing nipple 3/8"-1/2
- Compatible indicator for initial commissioning (TH2005, TH2025; TH2050; TH2100 or TH2250); in 500ml or 100ml)
- NeoTecMaster for multi-parameter integration
Certificates & approvals
- EN 61326-1
- EN 61010-1
- BS EN IEC 61326-1
- BS EN 61010-1+A1