These amperometric sensors with membrane are used to determine the concentration of free chlorine. They are able to measure various inorganic chlorine compounds, including chlorine gas (Cl₂), electrolytically generated chlorine, sodium hypochlorite (NaOCl), calcium hypochlorite (Ca(OCl)₂) and chlorinated lime (Ca(OCl)Cl).

Testing for the complete absence of free chlorine is not possible with these sensors.

Depending on the version, the sensors either supply a temperature-compensated current signal (4-20 mA) in an analogue version or a Modbus RTU signal via a digital interface. Calibration takes place in a downstream device such as an indicator, controller, recorder or programmable logic controller (PLC).

The sensors can be connected directly to various transmitters or controllers. These supply the system with the required voltage and enable uncomplicated calibration.

Areas of application:

The sensors are used to measure free chlorine in drinking water, bathing water, industrial water, process water and cooling water.

Dimensions

Height 8 cm
Length 30 cm
Width 20 cm

Weight

Weight 640 g

Other attributes

Country of origin DE
Customs tariff number 90278930
Technical data

Sensors for free chlorine

 

Sensor type

202630/40

202630/50

Measurand

Free chlorine

Area of application

Swimming pool water, drinking water, service water and process water

Suitable chlorination agents

Inorganic chlorine compounds: NaOCl (sodium hypochlorite), Ca(OCl)2, chlorine gas, chlorine pro- duced by membrane electrolysis (not suitable: chlorine electrolysis without a membrane)

Measuring principle

Membrane-covered, amperometric, two-electrode system with integrated electronics

Membrane type

Hydrophobic PTFE membrane

Measuring cable connection

2-pin terminal connection (2 × 1 mm2)

5-pin flange connector, M12

Voltage supply

UB DC 12 to 30 V

(galvanic isolation required)

UB DC 22.5 to 26 V

(galvanically isolated from the sensor)

Electromagnetic compatibilitya

Interference emission: class Bb Interference immunity: to industrial requirements

Output signal

4 to 20 mA

Modbus RTU

Load/current consumption

≤ (UB - 7.5 V) ÷ 0.02 A

approx. 20 mA

Settling time

1 h

Inflow speed

Approx. 15 cm/s

corresponds to a flow of approx. 30 l/h when installed in the JUMO flow fitting (part no.: 00392611)

Measuring rangesc

0.05 to 0.5 mg/l (ppm)

0.05 to 2 mg/l (ppm)

0.05 to 5 mg/l (ppm)

0.05 to 10 mg/l (ppm)

0.05 to 100 mg/l (ppm)

0.05 to 200 mg/l (ppm)

0.05 to 2 mg/l (ppm)

0.05 to 20 mg/l (ppm)

Resolution

0.01 mg/l with measuring range 0.5/2/5/10 mg/l

0.1 mg/l with measuring range 100/200 mg/l

0.001 mg/l with measuring range 2 mg/l

0.01 mg/l with measuring range 20 mg/l

Accuracyd

Measuring range 2 mg/l

 

< 1 % with 0,4 mg/l

< 1 % with 1,6 mg/l

Slope drifte

Approx. < -1 % per month

Response time t90

approx. 30 s

Operating temperature

Sample water temperature Ambient temperature

 

0 to 45 °Cf

0 to 55 °C

Temperature compensation

Automatic, using integrated temperature probe

Zero point adjustment

Not required

Slope adjustment

On evaluation unit/controller using analytical chlorine determination (DPD-1-method)

pH value operating range

pH 6 to pH 8

Note the effect of the pH value on the disinfecting properties, corrosion and the dissociation curve.

pH dependence (loss of slope)

with pH 8, approx. 65%

with pH 9, approx. 95% (starting at pH 7)

Disturbances

ClO2: recorded with a concentration factor of 9 O3: recorded

Chlorine electrolysis without a membrane can cause disturbances

Pressure resistanceg

Pabs max. 2 bar Prel max. 1 bar

Materials

Semi-permeable membrane, PVC-U

Dimensions

Dia. 25 mm, length 220 mm (housing with membrane cap)

Dia. 25 mm, length 205 mm (housing with membrane cap)

Weight

Approx. 125 g

a  EN 61326-1, EN 61326-2-3.

b  The product is suitable for industrial use as well as for households and small businesses.

c  Other measuring ranges upon request.

d  After calibration under replicable conditions (25 °C, pH 7.2 in drinking water) from the measuring range end value.

e  Under replicable conditions (25 °C, pH 7.2 in drinking water).

f  Prerequisite: no ice crystals in the measurement medium.

g  Pressure fluctuations are not admissible. Pressure-free operation (atmospheric pressure) recommended.

Skip product gallery

We recommend the following

JUMO single valve flow monitor, M12 process connection, close-up
JUMO single fitting flow monitor connection M12 process connection
Article number: 802153

Flow monitors for disinfection variables Membrane-covered sensors for water disinfection (e.g. free chlorine or total chlorine, ozone, chlorine dioxide, etc.) must be operated with a minimum flow rate of 15 cm/s. The flow monitor is integrated into the sample water supply line, in series with the disinfection sensor, and monitors the required minimum flow to the sensor. Below the minimum flow velocity, the sensors measure values that are too low. This can cause a connected control system to dangerously overdose or underdose. Above the minimum incident flow velocity, the measuring signal is only slightly influenced by the incident flow velocity. The flow monitor consists of a flow body, a needle valve insert for flow regulation and an inductive proximity switch, which is controlled by a float in the medium flow. When the minimum flow velocity is reached or exceeded, the float rises and the contact of the inductive proximity switch switches. The contact can be used, for example, to control a binary input of the controller for standard signals, JUMO AQUIS 500 AS. If the flow is too low, the controller is set to the ‘HOLD’ state. This prevents incorrect dosing.   Areas of application Flow monitoring when using sensors for free chlorine, chlorine dioxide, ozone, hydrogen peroxide or peracetic acid

€399.60*
JUMO ecoLine glass shaft electrode front view with ceramic diaphragm, ideal for water monitoring
JUMO ecoLine glass shaft pH electrode with KCL container, S8 Pg13.5 screw head
Article number: 800191
Diaphragm: Ceramic | Electrical connection of the electrode (onto which the cable is plugged): Screw head Pg13.5 (S8) | Extra code: KCL container (052) | Insertion length: 120mm | Measured parameter: pH | Measuring resistor (temperature sensor): without

The electrode measures in the pH range from 0 to 12, and for short periods up to pH 14. It is equipped with the low-ohmic JUMO UW glass, which supports fast and stable measurement results. JUMO ecoLine glass shaft pH electrode with KCL container, S8 Pg13.5 screw head – type key: 201005/51-18-22-0000-00-120/052, part no.: 00402273 This version of the JUMO ecoLine pH electrode is designed for reliable pH measurements in general aqueous media and offers a convincing price-performance ratio. Thanks to the additional KCL container, it is particularly practical when the sensor is not in continuous use, needs occasional refreshing, or is stored for longer periods. The ceramic diaphragm and polymerized KCl solid electrolyte make this sensor a suitable solution for many standard applications in water analysis. The proven JUMO reference system also helps ensure that the electrolyte remains silver-ion-free over its service life and is less susceptible to certain electrode poisons. Typical applications With S8 Pg13.5 screw head and 120 mm immersion length, this ecoLine pH electrode is suitable for many common installation situations. An integrated temperature sensor is not available in this series. The included KCL container is a useful advantage when the electrode is only used occasionally or needs to be stored safely between uses. drinking water monitoring water treatment swimming pools, aquaristics slightly polluted service, process and wastewater rainwater, pond and surface water applications horticulture applications food and beverage industry Not recommended for electroplating water treatment industrial process and wastewater ultrapure water therapy baths biotechnology and sterile processes

From €149.57*
JUMO ecoLine glass shaft pH electrode with salt reservoir, S8 Pg13.5 screw head, without temperature sensor, front view, suitable for drinking water monitoring
JUMO ecoLine glass shaft pH electrode with salt reserve, S8 Pg13.5 screw head
Article number: 800192
Diaphragm: Ceramic | Electrical connection of the electrode (onto which the cable is plugged): Screw head Pg13.5 (S8) | Extra code: Salt reservoir (837) | Insertion length: 120mm | Measured parameter: pH | Measuring resistor (temperature sensor): without

This electrode is particularly suitable wherever a cost-efficient, proven, and long-life-oriented electrode is required for general aqueous media. JUMO ecoLine glass shaft pH electrode with salt reservoir, S8 Pg13.5 screw head (type key: 201005/51-18-22-0000-00-120/837, part no.: 00447837) This JUMO ecoLine glass shaft pH electrode is a reliable choice for pH measurements in general aqueous media. It is designed for users who are looking for an economical, practical, and robust electrode suitable for many standard water applications. The version with salt reservoir is particularly advantageous when the electrode is used under conditions where extended service life is beneficial. Technical data Design: glass shaft pH electrode Measuring range: pH 0 to 12, short-term up to pH 14 Temperature sensor: none Active element: JUMO UW glass Diaphragm: ceramic Electrolyte: polymerized KCl solid electrolyte Reference system: JUMO cartridge reference system Electrical connection: S8 Pg13.5 screw head Immersion length: 120 mm (standard) Type extension: salt reservoir Advantages and special features The electrode measures in the pH range from 0 to 12, and for short periods up to pH 14. It is equipped with low-ohmic JUMO UW glass, enabling fast and stable measurement results. The ceramic diaphragm and polymerized KCl solid electrolyte make it suitable for many standard applications in water analysis. The proven JUMO cartridge reference system ensures that the electrolyte remains free of silver ions over its service life and is less susceptible to certain electrode poisons. Why choose this electrode? This version is particularly suitable when a pH electrode for general water applications is required that offers more than a basic standard model. The salt reservoir is especially beneficial in low-ion media or higher flow conditions, as it can extend the electrode’s service life. For this reason, it is ideal for users seeking a more durable solution for regular operation rather than a basic version without additional type extensions. With the S8 Pg13.5 screw head and 120 mm immersion length, the electrode fits many common installation situations. An integrated temperature sensor is not available in this series, keeping the focus entirely on reliable pH measurement. Typical applications drinking water monitoring and water treatment, swimming pools aquarium applications greenhouse and horticulture applications slightly polluted service, process, and wastewater rainwater, pond, and surface water use in portable measuring devices Not recommended for electroplating wastewater treatment industrial process and wastewater ultrapure water therapy pools biotechnology and sterile processes

From €149.07*
JUMO AQUIS RS Front view with illuminated LCD graphic display and plain text display, control panel with function keys and menu navigation
JUMO AQUIS RS, indicator/controller for digital sensors with Modbus protocol in surface-mounting housing, with RS485 Modbus master, analog output 0 (4) - 20mA or 0 - 10V output for main value
Article number: 800193
Unit selection: Version 1: Stock version: Type 202569/20-654-888-000-000-23/000 ; Part No. 00613453

The JUMO AQUIS RS is a practical solution when measurement values from digital sensors should not only be displayed but also directly processed and controlled. It is ideally suited for industrial applications in which measurement values are not only recorded but also evaluated and controlled. Digital sensors can be directly integrated via the integrated RS485 interface (Modbus). In addition, an analog temperature input is available, allowing external temperature sensors such as Pt100 or Pt1000 to be connected. The power supply for connected sensors is provided directly by the device, reducing installation effort. The LCD graphic display ensures a clear presentation of the measured values. Depending on requirements, these can be displayed as numeric values, bar graphs, or trends. The menu navigation is clearly structured and enables easy operation during ongoing use. In addition to the pure display function, the AQUIS RS also performs control tasks. With optional relay outputs, switching and alarm functions as well as more complex control functions with P, PI, PD, or PID behavior can be implemented. This makes the device suitable for both simple monitoring tasks and more demanding process control applications. Your advantages with the JUMO AQUIS RS Central display and processing of sensor data via Modbus RS485 Direct power supply of connected sensors via the device Large, easy-to-read LCD display with various display options Expandable for control and regulation tasks (including PID control) Versatile output signals for further processing (0–10 V, 4–20 mA) Possible application areas Environmental and laboratory measurement technology Water treatment systems Chemical and process engineering applications Applications with digital Modbus sensors Special features Display: mg/l, ppm, % SAT, %, ‰, g/l, ppb, μg/l, NTU, FNU, etc. Special display options are also possible with the setup program Large LCD graphic display with backlight Selectable display formats: large digits, bar graph, or trend display Integrated calibration routines: depending on the sensor: zero point, end value, and 2-point Calibration logbook Protection class IP67 for field mounting, IP65 for panel mounting Operating languages (switchable): German, English, French Via setup program (option): convenient programming, system documentation, loading additional operating languages Frequently Asked Questions What exactly can the JUMO AQUIS RS do? The device displays readings from digital sensors and can monitor and control them at the same time. Which sensors is the device compatible with? It works with digital sensors that support the Modbus protocol. Where can I use the device? For example, in water treatment, industrial plants, or anywhere where measured values need to be monitored. Can I transmit the measured values to other devices? Yes, values can be transmitted to other systems via the analogue output (0–20 mA, 4–20 mA or 0–10 V). Can I see the measured values directly on the device? Yes, the current values are displayed directly on the integrated display. Why is this device a good choice for my system? It combines display, monitoring and signal transmission in a single compact device, thereby simplifying process control.

€742.40*